Index:
[thread]
[date]
[subject]
[author]
From: eanders@uconnvm.uconn.edu
To : Joel Mackay , rasmb@alpha.bbri.org
Date: Tue, 10 Oct 2000 11:08:40 -0400
Absorbance OD instability
Joel,
We have seen this problem crop up a few times. It is a problem
with the software version and Firmware (EPROM) versions that you
are running. We solved this problem by returning to software
version XLI-3.01H and the correct EPROM's (DAB V 3.02 and ISB
V2.10) for this software version. Your Beckman Rep should be able
to deal with this problem.
Out of curiosity, which version of software and firmware do you
have?
Eric Anderson
NAUF
> Actually, while i am 'here', we also sometimes see the scans 'jump
> around', such that there are two distinct sets of overlapping scans in the
> one cell. I have explained this effect to myself by supposing that the
> monochromator is not selecting exactly the same wavelength each time it
> goes to that cell. The same questions as above apply here!
Index:
[thread]
[date]
[subject]
[author]